Occupied area: 1205*670*1686mm (L*W*H)
1Summarize:
SX-L1050 air filter media tester, our company developed high-end production in 1995, apply to test resistance of nonwoven filter material, glass fiber filter material or new type material. The whole system according to: EN779EN1822-3: 1998, IEST-RP-CC021.1, NIOSH, EN149: 2001, GB19083-2003 standard.
Recently, our clients all over China and abroad.
2Constituent parts:
1 filter;
2 atomizing;
3 flow equalize;
4 filter material installation;
5 test, control and data processing output;
6 blower power.
3Features:
1 Test area: 100cm2
2 filter speed: 5.33cm/s
3 Air volume: 32L/min
4 Aerosol section distribution: &10%
5 Aerosol particulate size: 0.3um(PSL)monodispersity
6 Upriver dilution: 7-10 times
7 Efficiency: 85%-99.995% @ 0.3um(concentration 80%)
Precise range:
Œ 99%-99.995% (Acceptable error after the last 9, for example: 99% acceptable range: 0.1%-0.9%, 99.95% acceptable range: 0.00%-0.09%; )
85%-99% (Acceptable error≤ ± 5%)
8 Traceable precise according to TSI3160 (Loaction: Tianjin national biological protective equipment and technology research center)
4Functions:
1 Test efficiency
2 Test resistance
5Major configuration and quantity:
1 PSL aerosol generator 1 set
2 750 W transducer 1 set
3 Diluter 1 set
4 Class 2.5 flowmeter 1 set
5 Particulate counter 2 set
6 pneumatic type filter material installation device 1 set
7 Data processing software 1 set
8 Self-purification 1 set
9 Electrical working desk 1 set
10Running tube(stainless steel) 1 set
11Static neutralizers 1 set
Sx-L1050 Air Filter Media Tester